- Design adheres to the principle of no moving parts, no moving
sample in sulfur recovery applications.
- The TLG-837 uses solid state diode array spectrophotometer
for detection, measuring complete spectrum from 190nm to 1100nm
with 1nm resolution
- The TLG-837 uses fiber optics to transmit light to and from
the detector, allowing the electronics to be physically separated
from the probe.
- a very long life pulsed xenon lamp or a low noise Deuterium
lamp. The light is focused onto the tip
- The analyzer measures a complete spectrum of the process and
analyze the data via a calibration method to give very accurate
and reliable concentration readings.
- Sampling - Patent-pending-cold-finger probe
- The inner one is basically a “cold finger” that
is designed to remove most of the sulfur vapor in a controlled
manner.
| Specification |
| Measurement method
: |
Diode array 190-1100
nm |
| Components
: |
H2S 0-2%
SO2 0-2%
COS 0-2,000ppm
CS2 0-2,000ppm |
| Accuracy
: |
±1%
of measurement on all components, (±5% below 500
ppm for COS and CS2) ±0.2% on air demand |
| Repeatability
: |
±0.4% of measurement
for H2S and SO2 ±0.1% for air demand. |
| Response
time : |
90% of final value in 10 sec. |
| Outputs
: |
4-20mA and RS232 for air demand,
H2S, SO2, COS, CS2 and sulfur vapor. |
| Area
Classification : |
Class1, Groups C & D,
Div. II /
Zone 1 II B T3 |
| Physical dimensions
(not including probe) |
| Size : |
30” H x 24” W
x 11” D (762mm
H x 610mm W x 280mm D) |
| Weight : |
164 lbs (74 Kg) |
|